1
This section covers the production of X-rays, stationary and rotating anode tubes, the control and indicating equipment of an X-ray machine, filters and grids, and the biological effects of X-rays.
2
Nature and Production of X-rays • X-rays are electromagnetic radiation of wavelength about 0.01-10 nm and energy of roughly 100 eV to 100 keV, discovered by Röntgen in 1895.
3
They are ionising, travel in straight lines at the speed of light, are undeflected by electric and magnetic fields, penetrate matter, cause fluorescence, blacken photographic film and produce biological effects.
4
The useful relation is E (keV) = 1.24 / λ (nm). • Production requires three things: a source of electrons (a heated tungsten filament, by thermionic emission), a means of accelerating them (a high potential difference, the kVp, between cathode and anode) and a target to stop them (the anode).
5
All of this takes place in an evacuated glass or metal envelope. • Two mechanisms of X-ray production: • Bremsstrahlung ('braking radiation') — an electron passing close to a tungsten nucleus is decelerated and deflected, and the lost kinetic energy appears as an X-ray photon.
6
Because the deceleration varies continuously, this gives a continuous spectrum from zero up to a maximum photon energy numerically equal to the applied kVp (an electron that loses all its energy in one interaction).
7
Bremsstrahlung accounts for the great majority of the beam in diagnostic radiology. • Characteristic radiation — an incident electron ejects an inner-shell (K-shell) electron of the target atom; an outer-shell electron falls into the vacancy and the energy difference is emitted as a photon of discrete energy characteristic of the target element (for tungsten the K-shell lines are at about 59 and 67 keV).
8
It therefore appears as spikes superimposed on the continuous spectrum, and only above the threshold kVp (about 70 kVp for tungsten). • Efficiency: only about 1 % of the electron energy becomes X-rays; about 99 % becomes heat — which governs the entire mechanical design of the tube.
9
Efficiency is proportional to Z × kV, which is why a high-atomic-number target (tungsten, Z = 74) is used.
10
The X-ray Tube • Cathode: a tungsten filament (usually two, for a fine and a broad focus) in a focusing cup of molybdenum or nickel, negatively charged so as to condense the electron stream onto the focal spot.
11
Filament current controls the temperature and hence, by thermionic emission, the tube current (mA), which is the number of electrons and therefore the quantity of X-rays. • Anode (target): tungsten or a tungsten-rhenium alloy on a molybdenum or graphite backing.
12
Tungsten is chosen for its high atomic number (74), very high melting point (3,370 °C) and good thermal conductivity.
13
In mammography a molybdenum or rhodium target is used instead, to give low-energy characteristic radiation suited to soft-tissue contrast. • The line-focus principle: the target face is angled (typically 6-17°) to the electron beam, so that the actual focal spot bombarded by electrons is large — spreading the heat — while the effective (apparent) focal spot projected towards the patient is small, giving good geometric resolution.
14
Effective focal spot = actual focal spot × sin θ.
15
The penalty is the anode heel effect: intensity is lower on the anode side of the field because those photons must traverse more target material, a difference of up to 45 % that is exploited by placing the thicker part of the patient towards the cathode. • Other components: the evacuated envelope (a vacuum is essential so that electrons are not scattered and the filament does not oxidise), the tube housing with lead shielding to absorb leakage radiation, the oil that provides insulation and heat transfer, the window and the collimator.
16
Stationary anode tube Rotating anode tube Construction A fixed tungsten button embedded in a copper block A tungsten-faced disc of 7-15 cm diameter on a molybdenum stem, driven by an induction motor whose rotor is inside the vacuum and stator outside Stationary anode tube Rotating anode tube Heat handling All the heat is delivered to one small spot; copper conducts it away The focal track sweeps a large circular area, spreading the heat over a track hundreds of times larger Speed — 3,000-3,600 rpm (standard) or 9,000-10,000 rpm (high speed) Tube rating Low — limited mA and short exposures Much higher — permits the high mA and short exposure times needed to freeze motion Focal spot Relatively large Can be small while still handling high loads, giving better resolution Cost and complexity Simple, cheap, robust Complex, expensive, bearing wear is the usual failure Typical use Dental units, portable and low-output machines All general radiography, fluoroscopy, angiography and CT • Heat units: for single-phase equipment HU = kVp × mA × s, multiplied by 1.35 for three-phase six-pulse and 1.41 for three-phase twelve-pulse or high-frequency generators.
17
Anode cooling charts and tube rating charts must not be exceeded. • Tube failure arises from filament evaporation and burnout, cracking or pitting of the anode surface, tungsten deposition on the glass envelope causing arcing, bearing failure, and loss of vacuum.
18
X-ray Control and Indicating Equipment • The generator supplies the tube: a high-tension (step-up) transformer producing the kilovoltage, a filament (step-down) transformer supplying a few volts at high current, rectifiers to make the supply unidirectional (the tube itself is a self-rectifier but this is inefficient), and the timer.
19
Generator types in ascending order of quality are single-phase half-wave and full-wave, three-phase six- and twelve-pulse, and high-frequency (constant potential) — the last being standard today because it gives an almost ripple-free output, higher mean photon energy, shorter exposure times and lower patient dose. • The three operator controls and what each does: kVp sets the accelerating potential and therefore the energy, penetration and beam quality; raising it increases both the quantity (roughly as kVp²) and the penetration, and reduces subject contrast. mA sets the tube current and therefore the number of photons.
20
Time (s) sets the duration; mA × s = mAs governs the total quantity of radiation, the film density and, directly and proportionally, the patient dose.
21
The classic exam rule is the 15 % rule — a 15 % increase in kVp approximately doubles the exposure, equivalent to doubling the mAs. • Automatic exposure control (AEC, phototimer) uses ionisation chambers or photomultiplier detectors behind the patient to terminate the exposure when the required radiation has reached the receptor, giving consistent density regardless of patient size.
22
A backup timer limits the maximum exposure. • Indicating and safety devices: kV and mA meters, exposure and ready indicators, the line voltage compensator, space charge compensation, the collimator with its light beam diaphragm (which must coincide with the radiation field within 2 % of the SID), filtration indicators, exposure switches that require the operator to be behind the barrier, interlocks, and the dose-area product meter. • Fluoroscopy adds an image intensifier (input phosphor of caesium iodide → photocathode → electrostatic focusing and acceleration to about 25-30 kV → output phosphor), giving brightness gain from flux gain and minification gain, with automatic brightness control and a 5-minute cumulative timer; flat-panel detectors have now largely replaced the intensifier.
23
Filters and Grids • Filtration removes the low-energy (soft) photons that would be absorbed in the patient's skin and contribute dose but not image.
24
Inherent filtration (the tube window, oil and housing, about 0.5-1.0 mm Al equivalent) plus added filtration (aluminium sheets) must total at least about 2.5 mm Al equivalent above 70 kVp.
25
The effect is beam hardening — the mean energy of the beam rises while the total intensity falls.
26
A compound (K-edge) filter such as erbium or the molybdenum filter of mammography selectively removes a band of energies; a wedge or compensating filter evens out exposure across a body part of varying thickness. • Attenuation obeys I = I₀ e−μx, where μ is the linear attenuation coefficient.
27
The half-value layer (HVL) — the thickness of a stated material reducing the intensity by half, HVL = 0.693/μ — is the standard practical measure of beam quality.
28
The interactions that matter diagnostically are the photoelectric effect (dominant at low energies and in high-Z materials, varying roughly as Z³/E³, and the source of most subject contrast and of patient dose) and Compton scattering (dominant above about 30 keV in soft tissue, nearly independent of Z, and the source of the scattered radiation that degrades contrast and endangers staff). • Grids are placed between patient and image receptor to absorb scattered radiation, which otherwise adds a general fog and destroys contrast.
29
A grid is an array of thin lead strips separated by a radiolucent interspace (aluminium, fibre or carbon). • Grid parameters: grid ratio = height of the lead strips ÷ width of the interspace (commonly 5:1 to 16:1 — a higher ratio removes more scatter but demands more accurate alignment and more dose); grid frequency (lines per cm); focused, parallel or crossed construction; and stationary or moving (the Potter-Bucky diaphragm, which blurs the grid lines out of the image).
30
Selectivity and contrast improvement factor describe performance, and the Bucky factor is the increase in exposure needed when the grid is used. • Grid cut-off — loss of primary radiation at the edges or across the image — occurs with off-level, off-centre, off-focus (wrong SID) or upside-down focused grid errors.
31
A grid is indicated when the body part exceeds about 10 cm or above about 70 kVp; alternatives are the air gap technique and tight collimation, which is always the first and best means of reducing scatter.
32
Biological Effects of X-rays • X-rays are ionising and act by direct damage to DNA and, more commonly, indirectly through radiolysis of water producing free radicals.
33
The critical target is DNA, and double-strand breaks are the lesions most likely to be lethal or mutagenic.
34
The law of Bergonié and Tribondeau states that cells are most radiosensitive when they are rapidly dividing, undifferentiated and have a long mitotic future — hence the high sensitivity of bone marrow, gonads, lymphocytes, intestinal crypt cells and the developing fetus, and the relative resistance of nerve and muscle. • Deterministic (tissue reaction) effects have a threshold dose, and their severity increases with dose: skin erythema, epilation, cataract, sterility, and the acute radiation syndromes.
35
They are the concern of fluoroscopy and interventional radiology. • Stochastic effects have no threshold; the probability (not the severity) increases with dose.
36
These are cancer induction and heritable genetic effects, and they are the reason for the linear-no-threshold model used in protection. • Units: exposure in coulomb per kilogram (old unit roentgen); absorbed dose in gray (Gy) = 1 J/kg (old unit rad, 1 Gy = 100 rad); equivalent dose in sievert (Sv) = absorbed dose × radiation weighting factor (1 for X-rays, photons and beta, 20 for alpha); and effective dose in sievert, weighting each organ by its tissue weighting factor to give whole-body risk. • Radiation protection rests on three ICRP principles — justification, optimisation (ALARA — as low as reasonably achievable) and dose limitation — and in practice on time, distance and shielding, distance being the most powerful because intensity obeys the inverse square law, I ∝ 1/d².
37
Practical measures: collimation to the area of interest, correct kVp and mAs, filtration, gonad and thyroid shielding, lead aprons (0.25-0.5 mm Pb) and lead glass, protective barriers and warning signs, personal dosimeters (film badge, TLD, OSL), the 10-day rule and pregnancy enquiry, and never holding the patient or cassette by hand.